Let's talk about your next study New study? Request a quote →
Loading...
Our Laboratory and Scientific Team

HIV-1 Recency Testing for Incidence Estimation

Limiting-antigen avidity EIA with reflex plasma viral load, for population-level HIV-1 incidence estimation

A positive HIV-1 antibody or nucleic acid result establishes that a person is infected. It does not establish when.

Once seroconversion is complete, a result from someone infected three months ago is indistinguishable from one from someone infected ten years ago, and for any question that turns on the rate at which new infections are occurring, rather than on their presence, that missing timing is the obstacle. The limiting-antigen avidity enzyme immunoassay (LAg-Avidity EIA), developed by investigators at the CDC to read that maturation, addresses this by measuring how far the antibody response has matured rather than whether it is present, so recent infection can be distinguished from long-standing infection in a single cross-sectional specimen. The result is reported as a normalized optical density (ODn) for each specimen, together with the recent or not-recent classification that the study’s algorithm assigns to it.

Reading recency from a single specimen, rather than from a cohort followed to seroconversion, lets a program estimate incidence in a population it has screened once. That makes a cross-sectional estimate available to a prevention trial that cannot ethically run a placebo arm, and to a survey that cannot bring participants back for a second visit.

What the Avidity Measurement Reads

After HIV-1 acquisition, antibodies against the virus become detectable within weeks, but their avidity, meaning the strength with which they bind antigen, continues to increase for months afterward as affinity maturation proceeds. That maturation is the signal the assay reads.

We run the assay under limiting antigen conditions, so the quantity of antigen available constrains binding, and we include a dissociation step that removes weakly bound antibody, which leaves a signal that depends on antibody avidity rather than on antibody concentration alone. We normalize each specimen’s optical density against a calibrator run on the same plate, which is what makes results comparable between plates and between runs, and that normalized value is the ODn. A low ODn is consistent with an immature, and therefore recent, antibody response, while a higher ODn reflects a matured response and long-standing infection.

Clinical and Preclinical Applications

We run the assay in two phases, screening and confirmatory.

01

In the screening phase we test each specimen once, and a specimen with a measured ODn above the initial screening threshold of 2.0 is reported as long-standing infection, while a specimen at or below that threshold proceeds to confirmatory testing.

02

In the confirmatory phase we retest each specimen in triplicate, and a specimen with a measured ODn at or below the confirmatory threshold of 1.5 is classified as recent infection, while one above 1.5 is reported as long-standing infection.

The antigen is a multi-subtype construct, which is why the assay has been applied across the major HIV-1 subtypes and circulating recombinant forms rather than only to subtype B populations.

Why the Assay Runs Inside an Algorithm

Several groups of people carry a low-avidity antibody profile without having acquired HIV recently, among them people on suppressive antiretroviral therapy, elite controllers who suppress without treatment, and some people with advanced untreated disease. Because recent infections are rare relative to prevalent ones in any screened population, even a small proportion of such false-recent results can dominate an incidence estimate.

Antiretroviral therapy can complicate recency classification, so people on ART are generally excluded from testing, and for those not identified as treated a reflex plasma viral load does the correcting. A specimen that confirms at or below the recency cutoff but carries a viral load below the reflex threshold is reclassified as long-standing infection, on the reasoning that a genuinely recent infection with suppressed viremia is rare. Independent evaluation of candidate incidence assays has reported high false-recent rates among virally suppressed people, and that is the misclassification the algorithm exists to control. A viral load criterion does not reach the false-recent results that arise in advanced untreated disease, where viremia is high, so those specimens remain in the false recency rate that the study’s statisticians apply when they adjust the estimate.

The avidity assay can be read on its own, but for incidence estimation the viral load is not an optional addition. The Forum for Collaborative Research consensus statement on recency testing in prevention trials describes the recent infection testing algorithm (RITA) as the avidity assay at an ODn threshold of 1.5 paired with a plasma viral load, and that pairing is what Accelevir recommends and what this service is built around. The ODn thresholds come from the assay and its calibrator rather than from a study-specific decision, so what a protocol settles in advance is the viral load threshold and which calibration parameters the analysis will use.

Counterfactual Incidence in Prevention Trials

Placebo-controlled trials of pre-exposure prophylaxis (PrEP) are no longer considered ethical, since effective prevention exists and must be offered to participants. Active-controlled designs are the alternative, but as the comparator becomes more efficacious the non-inferiority margin narrows and the required sample size grows, in some populations beyond what can feasibly be enrolled. For cisgender women, the consensus notes that efficacy against placebo in earlier trials of oral agents was inconsistent enough to preclude a defensible non-inferiority margin.

The Forum for Collaborative Research convened a working group of regulators, academic investigators, industry sponsors, statisticians, and community representatives to address this. Its 2023 consensus statement set out an alternative: estimate HIV incidence in people not on PrEP as an external counterfactual, using a recency algorithm applied to specimens from people identified as HIV-positive during trial screening, and compare on-PrEP incidence against that estimate.

That design is now more than a proposal, since the phase 3 lenacapavir trials used it. A recency algorithm applied to specimens from screening put background incidence at 2.41 per 100 person-years (95% CI 1.82 to 3.19) in PURPOSE 1 and at 2.37 per 100 person-years (95% CI 1.65 to 3.42) in PURPOSE 2. In each trial that estimate, rather than a placebo arm, was the comparator for the efficacy analysis.

The consequence for a sponsor is that screening-phase recency testing is no longer an ancillary measurement. It generates the comparator for the primary efficacy analysis, which puts the completeness of screening-phase specimen capture, the analytical performance of the testing laboratory, and the documentation behind the classification on the critical path of that analysis.

Two of the assumptions the consensus identified are study design questions rather than laboratory questions, and both are worth settling early because they determine what the laboratory needs to deliver and when. The first is that HIV acquisition risk in the people who test positive at screening is comparable to risk in the people who enroll, which is most readily achieved by applying eligibility criteria before the HIV test rather than after. The second is that willingness to be tested at screening is independent of HIV status, since people recently diagnosed elsewhere who do not present for screening are infections the estimate will not count.

HIV-1 Incidence Estimation in Vaccine Trials, Surveillance, and Cohort Studies

Preventive vaccine and long-acting agent trials.

The same counterfactual logic extends to preventive HIV vaccine candidates and to other long-acting agents, either as the primary comparator or as a supporting analysis alongside an active-control arm. Recency classification also bears on how a trial reads its breakthrough infections in aggregate, since the distribution of recent and long-standing results across them speaks to how many were likely acquired before product initiation rather than during it. That reading holds at the level of the set, not of any one participant.

Public health surveillance and program monitoring.

National surveys, key population surveys, and routine surveillance programs use algorithm-based cross-sectional incidence estimation to identify where transmission is concentrated and to assess whether prevention programs are changing incidence rather than only prevalence. Because the approach needs a single specimen rather than a follow-up visit, it can be built into survey designs that could not support a longitudinal incidence measurement.

Cohort and observational research.

Cohorts that enroll people already living with HIV can use recency classification at baseline to enrich for participants likely to have been infected recently, which is often the population of interest for studies of reservoir establishment, early treatment effects, and acute infection immunology.

What HIV-1 Recency Testing Includes

Both algorithm inputs from one laboratory.

The algorithm fixed in the study plan before testing begins.

Reporting built for the people who will use it.

Qualification against a recognized specimen panel, documented.

Capacity matched to screening, not to an average month.

Integration with the rest of an HIV-1 measurement program.

01

Both algorithm inputs from one laboratory.

An algorithm needs an avidity result and a quantitative viral load on the same specimen. We perform the LAg-Avidity EIA and the plasma HIV-1 viral load in the same facility, the viral load as a single standard measurement on the Hologic Aptima HIV-1 Quant Dx assay on the Panther system rather than the multi-replicate ultrasensitive assay, which answers a different question. Generating both inputs here means one accessioning record, one chain of custody, and no second shipment of aliquots to a separate viral load vendor.

02

The algorithm fixed in the study plan before testing begins.

We record the reflex viral load threshold and the calibration parameters in the study-specific plan and apply them as written, because post hoc changes to an algorithm are difficult to defend in an efficacy analysis.

03

Reporting built for the people who will use it.

Deliverables include the specimen-level ODn with run-level control performance, the paired viral load result, and the classification each specimen received under the agreed algorithm, formatted so that your statisticians can apply an incidence estimator directly. We report the classification logic as applied, so the analysis is reproducible from the raw data.

04

Qualification against a recognized specimen panel, documented.

We run recency testing as an analytically qualified assay under Accelevir standard operating procedures and controls, and we qualified it here against a 99-specimen panel from the CEPHIA repository, the specimen collection assembled for independent evaluation of HIV incidence assays, drawn from people living with HIV in five countries and covering subtypes B and C. The qualification covered intra-assay, inter-assay, and inter-operator precision, operator bias, and the reproducibility of recency classification for specimens sitting on either side of the cutoff, each against criteria set before the work began. The qualification report is available on request. The laboratory also enrolled in the EQAPOL limiting-antigen avidity proficiency program, run with the CDC and the Blood Systems Research Institute, and passed proficiency testing. We take part in proficiency rounds while recency work is active, and re-enroll for a study that calls for it.

05

Capacity matched to screening, not to an average month.

Prevention trials generate recency specimens in concentrated bursts during screening, on a timeline that is hard to predict in advance. We scope batch capacity and turnaround against the enrollment plan rather than against a monthly average.

06

Integration with the rest of an HIV-1 measurement program.

Recency classification sits alongside the reservoir and virologic assays we run in the same laboratory, so a program characterizing breakthrough infections can add HIV-1 IPDA®, cell-associated HIV-1 RNA, and standard or ultrasensitive plasma viral load to the same specimen set.

Specimen Requirements for HIV-1 Recency Testing

Plasma is the specimen type for recency testing, because the algorithm’s viral load component has to be quantified from the same specimen and the Aptima assay cannot quantify viral load from serum. Dried blood spots have been used for avidity testing in surveillance settings, but the false-recent rate is higher and the ODn less stable than from plasma, and dried blood spot algorithms carry their own calibration parameters, so we recommend plasma and discuss dried blood spots only where a program has no practical alternative. Approximately 1 mL of plasma is a typical starting requirement, covering the avidity assay, confirmatory testing where the algorithm calls for it, and the reflex viral load. We set requirements with you per project where less specimen is available.

Negative selection

Several design decisions influence the final incidence estimate more than the assay does, and we raise them early:

The screening population and the enrolled population.

The counterfactual estimate is interpretable only if the people contributing to it resemble those who enroll in acquisition risk. Site selection, recruitment method, and the order in which eligibility and HIV testing happen all bear on this.

Treatment status at screening.

The laboratory cannot identify a treated participant from the specimen alone, so capturing treatment status at screening is what allows those specimens to be excluded.

Prior and concurrent PrEP use.

Antiretroviral exposure around the time of acquisition may alter antibody maturation and therefore the classification. The consensus recommends excluding people already on PrEP from the cohort used to generate the counterfactual estimate, or handling them in a sensitivity analysis, so capturing prior PrEP use at screening is worth building into the case report form.

Subtype distribution.

Calibration parameters derived from one subtype distribution do not transfer cleanly to another. The population’s subtype composition should inform which values your statisticians apply, and it usually comes from molecular epidemiology data for the region. Accelevir scientists can discuss HIV-1 RNA sequencing of the study specimens at scoping where a program would rather determine subtype directly.

Specimen volume and retention.

Confirmatory testing and any repeat testing under the same algorithm come from one collection, so aliquoting should anticipate the full algorithm rather than the first test in it.

What Recency Testing Does and Does Not Capture

What it captures

What it does not capture

An individual diagnosis or date of acquisition.

Classification error at the individual level is substantial. A recent classification is not a clinical finding, should not be returned to or about a single participant, and should not be used to infer when one person acquired HIV. Study protocols and informed consent documents should be explicit on this point.

The incidence estimate itself.

We generate the algorithm inputs and the classification, and converting them into an incidence estimate requires the mean duration of recent infection and the false-recent rate appropriate to the population, so that calculation and its adjustment are performed by your statisticians rather than here.

A result independent of external calibration.

The accuracy of any estimate built from these data depends on calibration parameters that are properties of the algorithm and the population rather than of the specimen, which is why we record them in the study-specific plan before any specimen is tested.

Reservoir size or persistence.

Recency describes time since acquisition, while reservoir questions are answered by HIV-1 IPDA®, cell-associated HIV-1 RNA, and ultrasensitive plasma viral load.

Our Laboratory and Scientific Team

About Accelevir

We design assays around signals that are uncommon in the population being tested, and a recent infection among people screening HIV-positive is one of them: the classification that carries the estimate is the minority one, and the cost of getting it wrong lands on the incidence figure rather than on the plate. Accelevir began as a Johns Hopkins spinout focused on detecting the rare, intact HIV-1 that persists in people on therapy, and that work shaped how we approach measurements of this kind. We have supported recency testing for an HIV prevention trial, and Accelevir has supported 100+ clinical studies, of which dozens are interventional cure and remission trials.

Qualification against an independent repository panel, with reporting your statisticians can use directly. We qualified the assay against a CEPHIA panel rather than against our own material alone, and we report the ODn, the paired viral load, and the classification in the form the standard incidence estimators take, so the analysis can be reproduced from the raw data.

One chain of custody for a classification that rests on two measurements. Because both algorithm inputs are generated here, the recency call for each specimen carries one accessioning record and one custody trail rather than results assembled across vendors.

All samples are processed in our BSL-2+ laboratory, which adds enhanced biosafety practices to BSL-2 containment for work with bloodborne pathogens and viral vectors. We run recency testing as an analytically qualified assay under the quality management system that underpins our laboratory accreditation, with qualification documentation available to support regulatory filings, and the recency classification is a research and surveillance output rather than a diagnostic result. Accelevir is scientist run, with one team from protocol design to final report, so a study design can adapt as the program develops.

References

FAQ

Frequently Asked Questions

How does a recency assay differ from a fourth-generation HIV test or a nucleic acid test?

Fourth-generation antigen and antibody tests and nucleic acid tests are designed to shorten the interval between acquisition and detectable infection, so they identify infection earlier. They still classify status rather than timing, and once the result is positive it does not distinguish a recent acquisition from an established one. Recency testing works in the opposite direction on the same timeline: it is applied only to specimens already known to be HIV-positive, and asks how long the infection has been present.

Can a recency result be given to an individual participant?

No. Classification error at the individual level is substantial enough that these results support population-level incidence estimation only. They should not be reported as a clinical finding or used to infer an individual’s date of acquisition.

Do you calculate the incidence estimate?

No. We report the specimen-level ODn, the paired viral load, and the classification under the algorithm agreed in the study plan, formatted for your statistical team. The incidence estimate and its adjustment for the mean duration of recent infection and the false-recent rate are performed by your statisticians, who hold the calibration parameters appropriate to the study population.

Why does the algorithm require a viral load result?

People with suppressed viremia, whether from antiretroviral therapy or from natural control, frequently produce low-avidity results that would otherwise be classified as recent. Requiring a viral load above a defined threshold removes most of that misclassification before any estimate is calculated.

Does the assay work across HIV-1 subtypes?

The assay reads a multi-subtype antigen and has been applied across the major subtypes and circulating recombinant forms. Performance is not identical across all of them, which is why subtype-appropriate calibration parameters, rather than a single pooled value, should be used in the estimate.

What specimen do you need?

Plasma, with approximately 1 mL as a typical starting requirement to cover the avidity assay, confirmatory testing, and the reflex viral load. Serum is not suitable, because the viral load component of the algorithm cannot be quantified from serum on the Aptima assay. Dried blood spots are used in some surveillance settings, but the false-recent rate is higher and the ODn less stable than from plasma, so we recommend plasma.

Can recency testing substitute for a prospective cohort?

It addresses a different constraint rather than replacing cohort measurement. Where a prospective cohort is feasible, it remains the more direct measurement of incidence. Cross-sectional estimation is the practical option where a placebo or untreated comparison group is not ethically available, or where a survey cannot support follow-up, and the consensus work in prevention trial design reflects that reasoning rather than a claim of equivalence.

What turnaround should we expect?

Turnaround depends on batch size, on whether the reflex viral load runs concurrently, and on the reporting format agreed in the study plan. We scope it against your screening timeline during study setup.

Is recency testing available for programs outside clinical trials?

Yes. Surveillance programs, key population surveys, and academic cohorts are in scope, and we set the reflex viral load threshold and the reporting format against the program’s analysis plan in the same way.

Match Recency Testing to Your Incidence Question

Complete the project form with the incidence question and the population, the screening or survey timeline, the subtype distribution you expect, the specimen type and volume available, and what your statisticians need back. A scientist will propose the testing plan, the specimen requirements, and the batch schedule, and return an itemized quote matched to your study design.

Accelevir Diagnostics, LLC is a CAP-accredited, CLIA-certified laboratory. Services are provided for research use in support of clinical and preclinical studies and are not intended for diagnostic, therapeutic, prognostic, or patient-management use.