A Scientist’s Guide to Immunogenicity Testing in Biologics

NAb Assay Development
for Biologics, Cell Therapies, and Gene Therapies

We run NAb assays in three formats: a plate-based competitive ligand-binding format, a cell-based format with an ELISA readout, and a flow cytometry format.

A positive anti-drug antibody (ADA) result shows that antibodies to a drug have formed, not what they do. Some recognize an epitope that leaves the drug’s interaction with its target or receptor intact; these non-neutralizing antibodies do not block the drug’s activity directly, although they may still affect its exposure. Others bind in a way that prevents that interaction, and these neutralizing antibodies (NAbs) can reduce or abolish the drug’s activity. When your program needs to know which kind of response is present, a NAb assay answers that question.

Ligand-Binding, Cell-Based, and Flow Cytometry NAb Formats

Plate-based ligand-binding format

Cell-based format with an ELISA readout

Flow cytometry format

Choosing a format

01

Plate-based ligand-binding format

This competitive format measures whether antibodies in the sample block the drug from binding its purified receptor or target ligand on a plate. It is faster and less expensive to run than a cell-based format, is often suited to Phase 1 clinical trials, and requires the purified receptor or ligand.

02

Cell-based format with an ELISA readout

A cell line that overexpresses the drug’s receptor and responds with a measurable output, such as IL-2 production, is cultured with a known concentration of drug and the study sample. Neutralizing antibodies keep the drug from acting on the cells, so the reporter signal measured by ELISA falls as neutralizing activity rises.

03

Flow cytometry format

Cells that overexpress the drug’s receptor are incubated with fluorochrome-conjugated drug and the study sample. Neutralizing antibodies keep the drug from binding the cells, so median fluorescence intensity falls as neutralizing activity rises.

04

Choosing a format

The plate-based and flow cytometry formats measure whether antibodies block the drug from binding its receptor, on a plate or on cells, while the ELISA-readout format measures whether they block the cellular response the drug produces. The choice follows the drug’s mechanism of action and the program’s stage, and we work through it with you at scoping.

Where NAb Testing Sits in an ADA Workflow

NAb testing follows the ADA tiers. Samples are screened, putative positives are confirmed, and confirmed positives are titered and, where the program calls for it, tested for neutralizing activity. Drug-tolerant pretreatments, such as acid dissociation, address circulating drug in NAb assays as well as in the ADA tiers.

Screening

Presence of binding antibody

Confirmatory

Binding is specific

Titer

Magnitude and persistence of the response

Neutralizing antibody

Drug function is blocked

When Should Sponsors Start NAb Assay Development for a New Biologic?

Start before the first study whose samples will be tested for neutralizing activity. NAb testing is generally expected in clinical trials and optional in many preclinical studies. For a regulated study, the method is validated first, and sample analysis begins once you have approved the validation results. Settling the format, the positive control, and the drug-tolerance approach during preclinical work lets those choices carry into the clinical method.

Development is run as non-GLP work. NAb assays can be qualified for non-GLP studies and validated according to current FDA guidance for regulated preclinical and clinical studies, under GLP when required. Validation covers parameters such as the cut point, sensitivity, precision, selectivity, drug tolerance, and stability. We run NAb testing as a standalone service or within a tiered immunogenicity plan alongside our ADA assays, and when your program needs preclinical in-life work, we can support it through collaborating laboratories.

Preclinical work:format, positive control,drug-tolerance approachClinicalmethodValidation resultsapprovedSample analysisbegins
CAR-T programsChimericantigen receptorGene therapyprogramsVector andtransgene product

What Does NAb Work Involve for Gene Therapy and Cell Therapy Programs?

Neutralizing antibodies can arise against cell and gene therapies as well as protein biologics. In CAR-T programs, they can form against the chimeric antigen receptor, and in gene therapy programs, against the vector and against the transgene product. For gene therapy programs, we also measure vector biodistribution and shedding by digital PCR.

What a NAb Readout Does and Does Not Capture

What a NAb result reports

A NAb result reports whether antibodies in a confirmed-positive sample inhibit the drug's activity, or its binding to its receptor, in a defined in vitro assay system.

01

Antibodies that bind without inhibiting.

It does not detect antibodies that bind the drug without inhibiting it, which the screening and confirmatory ADA tiers detect.

02

What happens in a study participant.

It does not show what happens in a study participant, because inhibition in the test system is not a measure of clinical consequence.

03

A single sensitivity figure.

Sensitivity and drug tolerance are set during development for each drug and matrix, so they are reported for each assay rather than as a single figure.

Start Your Project

If your biologic, gene therapy, or cell therapy program needs a neutralizing antibody assay, bring us your drug, the species and sample types involved, and the question your next milestone depends on. Our scientists will scope the assay design, validation path, and sample plan with you and return an itemized quote.
FAQ

Frequently Asked Questions

Which sample types can we send for ADA and NAb testing?
ADA and NAb assays are run on serum or plasma, not on cells. We test samples from human and nonclinical species, and we support collection protocols with timepoint-based sampling. We do not quote a single volume, because needs vary with the assay, matrix, and species. Tell us the drug, the species, and your collection schedule, and our scientists will confirm what to send and how to ship it.
How is sample storage handled across a long immunogenicity study?
We manage long-term sample storage with chain-of-custody integrity, which may help keep longitudinal timepoints traceable. Planning timepoint-based sampling at the start helps serum and plasma arrive consistently.
What does study coordination and data reporting look like for an ADA and NAb program?
We offer study coordination and high-throughput sample handling alongside ADA and NAb work, and one team follows the study from protocol design to final report. Bioinformatics and statistical analysis are offered for reporting, so titer and functional results can be read next to your other endpoints.
Can a NAb assay be read out by flow cytometry?
Yes. Alongside the plate-based ligand-binding format and the cell-based format with an ELISA readout, we offer a flow cytometry format, in which neutralizing antibodies reduce the binding of fluorochrome-conjugated drug to cells that overexpress its receptor.
How do we start scoping an ADA and NAb study with your team?

Send the project form to talk through your molecule, sample matrices, and timepoints with our scientists. Our scientific advisory and consulting also extends to biomarker planning, QA/QC oversight, client onboarding, and scientific data interpretation. Bringing your sampling plan and endpoints to that first conversation lets us match assay formats to your question.

Can ADA and NAb methods be transferred or bridged between laboratories?
Yes. If your program changes laboratories or stages, our scientists can bring an existing ADA or NAb method into our workflow and check that it behaves consistently, so earlier timepoints stay comparable with later ones. Share the current method documents and sample plan in your first conversation, and we will scope the transfer.